EASYBONE Quattro

Surgical micromotor for implant physiodispenser

Suitable for implants, and maxillo-facial surgery and microsurgery procedures.

  • BRUSHLESS MIRCROMOTOR with 200 to 40.000 rpm (1:1) speed range: simple and pratical functioning at every speed/torque option, and selected program;
  • MICROMOTOR with contra-agle anti-rotation device. Stainless steel mototr case and connection cable can be both subject to severat auto-cleavable;
  • TORQUE from 01 to 80 Ncm with contrangle 32:1;
  • SPEED with contrangle 32:1 from 6 to 1250 rpm
  • 33 SETTABLE REDUCTIONS;
  • PUMP IRRIGATION LIQUID FLOW adjustable from 0 to 100%;
  • DIRECT ACCESS PROGRAMMABLE FUNCTIONS: speed, torque, pump, reverse, reducer hand-piece (all parameters in sequential mode);
  • ERRORS – DIAGNOSTIC SYSTEM CONTROL;
  • ANGLE AUTO CALIBRATION upon machine’s activation;
  • DIRECT SWITCH FOR EVENTUAL PUMP’S ACTIVATION;
  • INSTANT VISUALIZATION OF TORQUE AND SPEED ADJUSTEMENTS, during the operative phase;
  • POSSIBILITY TO CHANGE THE OPERATION PARAMETERS and opportunity to memorize them;
  • 4 PRE-PROGRAMMED FUNCTIONS by direct button.

Easybone DI 500 with Light

  • LED Light high efficiency,
  • higher visibility of the treated area
  • higher precision

Equipped with:

  • electronic control with peristaltic pump
  • high-speed motor with lighting
  • multifunction pedal DI 500PV (optional)
  • ON/OFF pedal SB 400P (optional)
  • contrangle with light model CIS321SNLIGHT (optional)

Easybone DI 400 without light

Equipped with:

  • electronic control with peristaltic pump
  • high-speed motor
  • multifunction pedal DI 500PV (optional)
  • ON/OFF pedal SB 400P (optional)
  • contrangle 32:1 model CI321SN (optional)

Display

The visualized parameters are specified as follows.
Please make reference to the above figure:

A – it indicates the selected function (P1, P2, P3, P4)
B – it indicates the speed
C – it indicates the peristaltic pump power’s percentage
D – it indicates the torque
E – it indicates the reduction/increase of the selected parameters

UP – By pushing this button the selected blinking parameter (speed, torque, reduction or pump) can be increased. DOWNBy pushing this button the selected blinking parameter (speed, torque, reduction or pump) can be decreased.

SET

By pushing this button, the parameter (speed, torque, reduction or pump), can be sequentially selected, then when the desired parameter is blinking. It can be adjusted.

CAL

Insert the contra-angle with the bur. Press the button CAL and the pedal until calibration OK.

REVERSE

By pushing this button, the reverse device can be temporary activated or deactivated. When the pedal is released, the forward mode is reintroduced.

MEMORY

By pushing this button, the selected blinking parameter (speed, torque, reduction or pump), can be memorized.

PUMP

By pushing this button, the pump status can be modified (from ON to OFF, or vice versa).

Optional

Multifuctional pedal control (optional) with reverse direction. on-off pump, start speed regulation and possibility to change functions PI/P2/P3/P4. The pedal controls Micromotor start. Suitable for operating theatres and sterile environments.

Pneumatic on-off footswitch (optional) suitable for operating theatres and sterile environments.

  • Contra-angle 20:1model CI 201MX without light
  • Contra-angle 20:1model CI 201MXLIGHT with light
  • High torque resistant gears
  • Push button
  • Easy to disassemble and lubricate
  • DI IMS: irrigation pipe

Technical features

  • Power source: 230V+/-10% | 50/60 Hz                                                    100-115V 50/60 Hz (optional)
  • Nominal power consumption: 170 VA
  • RPM: 200 MIN – 40.000 MAX CON RID.1/1
  • Hydraulic circuit capacity: MAX 50 ML/MIN
  • Pneumatic or multifunctional footswitch
  • Weight: 4,1 Kg
  • Dimensions: WxDxH: cm 25,5x25x12,5
  • DI IMS – irrigation pipe
  • Speed
    Instant visualization of torque and speed adjustments, during the operative phase.
  • Auto calibration
    Contra-angle autocalibration upon machine’s activation.